Table 9: Comparison of experimental and simulated coincidence summing correction factors of 152Eu for Marinelli beaker source.

Volumes (ml) Energy (keV) CSFExp CSFCal J 1 CSFSimu RD (%)
450 778.9 0.870 0.887 0.111 0.889 -2.0
964.0 0.894 0.922 0.922 0.912 -2.0
444.0 0.888 0.859 0.095 0.905 -1.9
600 778.9 0.967 0.893 0.030 0.970 -0.3
964.0 0.899 0.927 0.085 0.915 -1.7
444.0 0.894 0.867 0.088 0.912 -2.0
800 778.9 0.870 0.896 0.110 0.890 -2.2
964.0 0.895 0.929 0.090 0.910 -1.6
444.0 0.894 0.870 0.095 0.905 -1.2
1000 778.9 0.875 0.897 0.108 0.892 -1.9
964.0 0.904 0.930 0.087 0.913 -0.9
444.0 0.908 0.972 0.083 0.917 -0.9