Table 3: Comparison between experimental and simulated coincidence summing correction factors for the cylindrical source.
Volume (ml) | Energy (keV) | CSFExp | CSFCal | CSFSimu | RD (%) |
50 | 1173.24 | 0.917 | 0.923 | 0.924 | -0.7 |
1332.50 | 0.916 | 0.921 | 0.922 | -0.6 | |
898.02 | 0.957 | ||||
1836.01 | 0.901 | ||||
100 | 1173.24 | 0.921 | 0.927 | 0.926 | -0.5 |
1332.50 | 0.905 | 0.925 | 0.925 | -2.1 | |
898.02 | 0.951 | ||||
1836.01 | 0.892 | ||||
200 | 1173.24 | 0.923 | 0.935 | 0.927 | -0.4 |
1332.50 | 0.911 | 0.934 | 0.925 | -1.5 | |
898.02 | 0.942 | ||||
1836.01 | 0.880 | ||||
300 | 1173.24 | 0.922 | 0.938 | 0.928 | -0.6 |
1332.50 | 0.905 | 0.937 | 0.926 | -2.2 | |
898.02 | 0.938 | ||||
1836.01 | 0.872 |