Table 1: Injector program.
Line |
Function |
Amount |
Vial position |
Reagent |
1 |
Draw |
2.5 μL |
1 |
borate buffer |
2 |
Draw |
0.5μL |
7 onwards* |
sample |
3 |
Mix |
3.0 in air max speed 2 times |
|
|
4 |
Wait |
0.5 min |
|
|
5 |
Draw |
0 μL |
2 |
Water - (needle wash using uncapped vial) |
6 |
Draw |
0.5 μL |
3 |
OPA |
7 |
Mix |
3.5 μL in air Max speed 6 times |
|
|
8 |
Draw |
0.5 μL |
4 |
FMOC |
9 |
Mix |
4.0 μL in air Max speed 6 times |
|
|
10 |
Draw |
0 μL |
2 |
Water- (needle wash using water in uncapped vial) |
11 |
Draw |
|
6 |
acetonitrile |
12 |
Draw |
32 μL |
5 |
water (capped vial) |
13 |
Mix |
18 μL in air max speed 2 times |
|
|
14 |
Inject |
|
|
|
*vial position of the sample in the tray.