Table 2B: Between-batch precision and accuracy of the method for determining DAS in plasma samples.

Batch No: 1-3 (n = 18)
Nominal Conc. (ng/mL) Conc. Found (mean ± SD; ng/mL) RE (%) CV (%)
1 1.02 ± 0.04 1.59 4.21
3 3.05 ± 0.07 1.53 2.34
40 40.02 ± 0.38 0.04 0.94
160 164.99 ± 2.30 3.12 1.34
320 326.84 ± 5.40 2.14 1.65

Conc: Concentration; n: Replicates at each concentrations; RE: Relative Error; CV: Coefficient of Variation; SD: Standard Deviation.